• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL ENSITE¿ PRECISION¿ LINK SENSOR ENABLED; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL ENSITE¿ PRECISION¿ LINK SENSOR ENABLED; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number H702475
Medical Device Problem Code Arcing (2583)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
This report is to advise of an event observed during analysis confirming two discolored/blackened pins which was isolated to an electrical arc at the rear power inlet connector.
 
Additional Manufacturer Narrative
One ensite precision link sensor enabled was received for analysis with no accessories or original packaging available for inspection.Visual inspection of the returned device revealed the fiber com board was observed to have two discolored/blackened pins.The two discolored/blackened pins was isolated to an electrical arc at the rear power inlet connector.Root cause of the observed damage remains undetermined.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSITE¿ PRECISION¿ LINK SENSOR ENABLED
Common Device Name
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key13591638
Report Number2184149-2022-00054
Device Sequence Number14907871
Product Code DQK
UDI-Device Identifier05415067024756
UDI-Public05415067024756
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K183128
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberH702475
Device Catalogue NumberH702475
Device Lot Number6344643
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/18/2022
Initial Report FDA Received Date02/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/02/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-