MAUDE Adverse Event Report: SMITH & NEPHEW, INC. OXINIUM FEM HD 12/14 36 MM M/+4; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. OXINIUM FEM HD 12/14 36 MM M/+4; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 71343604
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 02/04/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that, after thr surgery had been performed on (b)(6) 2022 the patient experienced dislocation due to the components being placed in a sub-optimal position.This adverse event was solved by revision surgery to explant the complete system on (b)(6) 2022.The physician revised an oxinium fem hd 12/14 36 mm m/+4.The current health status of the patient is unknown.
 
Additional Manufacturer Narrative
The device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, he patient had a thr with the r3 system and two days later had a revision due to dislocation.It was noted via the attachments that no further clinical documentation is available.Therefore based on insufficient information, a thorough medical investigation cannot be performed.However, we are unable to rule out a procedural variance as a contributing factor to the reported dislocation as it was noted that the components were placed in a suboptimal position.The surgical technique states that the final position of the acetabular component should be approximately 45 degrees of abduction and 20 degrees of anteversion or at the surgeon¿s desired acetabular cup orientation.No further medical assessment can be rendered at this time.Should additional clinical documentation become available in the future, the clinical/medical task may be re-evaluated.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A review of the instructions for use for this device reveal in the warnings and precautions, intraoperative section, that proper positioning of the components is important to minimize impingement which could lead to early failure, premature wear, device related noise, and/or dislocation, all of which may lead to revision surgery.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to traumatic injury, patient anatomy or abnormal loading of limb.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXINIUM FEM HD 12/14 36 MM M/+4
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13591688
Report Number1020279-2022-00853
Device Sequence Number20320917
Product Code JDI
UDI-Device Identifier03596010477286
UDI-Public03596010477286
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K022958
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71343604
Device Catalogue Number71343604
Device Lot Number20HM02148
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/04/2022
Supplement Date Received by Manufacturer03/24/2022
Initial Report FDA Received Date02/24/2022
Supplement Report FDA Received Date03/25/2022
Date Device Manufactured08/05/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-