• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PFC*SIGMA DI AUG16MMCO,SZ3LEFT; SIGMA REVISION IMPLANT : KNEE FEMORAL ACCESSORY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PFC*SIGMA DI AUG16MMCO,SZ3LEFT; SIGMA REVISION IMPLANT : KNEE FEMORAL ACCESSORY Back to Search Results
Model Number 96-0871
Medical Device Problem Code Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/15/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that a cast product cleanroom (cpcr) 2nd shift finish operator was performing the wi-0897 ¿repackaging / resterilization of sterile product in raynham¿ for a pfc sigma augment repackaging order.During this process, the operator identified multiple parts displayed a visual cosmetic defect on the top surface.The operator notified the finish lead, who contacted process and quality engineering for further evaluation.The image attached shows a augment returned from expiry with foam particles transferred from the packaging components.This pc was created to capture the products that are also involve in the complaint reported in main (b)(4).
 
Additional Manufacturer Narrative
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information, which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained, that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary examination of the returned sigma augment confirmed inner foam packaging residue/debris was found transferred to the surface of the device.The debris was found in raynham during repacking/resterilization of expired sterile product returned from the field.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the product and lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
SIGMA REVISION IMPLANT : KNEE FEMORAL ACCESSORY
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY RAYNHAM, A DIV. OF DEPUY ORTHO 1219655
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key13592367
Report Number1818910-2022-03477
Device Sequence Number3860850
Product Code JWH
UDI-Device Identifier10603295234340
UDI-Public10603295234340
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K952830
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number96-0871
Device Catalogue Number960871
Device Lot NumberE22MF4
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/31/2022
Supplement Date Received by Manufacturer04/26/2022
05/25/2022
Initial Report FDA Received Date02/24/2022
Supplement Report FDA Received Date04/26/2022
05/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/28/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PFC SIGMA FEM POST AUG SZ2 4MM; PFC SIGMA FEM POST AUG SZ2 4MM; PFC SIGMA FEM POST AUG SZ2 4MM; PFC SIGMA FEM POST AUG SZ2 4MM; PFC SIGMA FEM POST AUG SZ2 4MM; PFC*SIGMA DI AUG16MMCO,SZ3LEFT; PFC*SIGMA DI AUG16MMCO,SZ3LEFT; PFC*SIGMA DI AUG16MMCO,SZ3LEFT; PFC*SIGMA DI AUG16MMCO,SZ3LEFT; PFC*SIGMA DI AUG16MMCO,SZ3LEFT; PFC*SIGMA DI AUG16MMCO,SZ3LEFT; PFC*SIGMA DI AUG16MMCO,SZ3LEFT; PFC*SIGMA DI AUG16MMCO,SZ3LEFT; PFC*SIGMA DI AUG16MMCO,SZ3LEFT; PFC*SIGMA DIS AUG4MM,S2.5,RGHT; PFC*SIGMA DIS AUG4MM,S2.5,RGHT; PFC*SIGMA DIS AUG4MM,S2.5,RGHT; PFC*SIGMA DIS AUG4MM,S2.5,RGHT; PFC*SIGMA DIS AUG8MM,S2.5,LEFT; PFC*SIGMA DIS AUG8MM,S2.5,LEFT
Patient SexUnknown
-
-