The device was returned for evaluation.Upon evaluation of the device, the reported issue was confirmed.An industrial scope was used to check the inner wall of the ch-tube and black material on the inner wall from the distal end to the 30mm depth of the ch-tube was observed.In addition, no fluid invasion was found inside the lg-cover or connector, no leakage occurred at leakage test, no anomaly was found at image check, no defect was found on the eyepiece or connector, and no anomaly was found on the insertion section or bending section.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time, however, if additional information becomes available, this report will be supplemented accordingly.
|