| Model Number |
96-0911 |
| Medical Device Problem Code |
Tear, Rip or Hole in Device Packaging (2385)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/31/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # pc-(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2022, a cast product cleanroom (cpcr) 2nd shift finish operator was performing the wi-0897 ¿repackaging / resterilization of sterile product in raynham¿ for a pfc sigma augment repackaging order.During this process, the operator identified multiple parts displayed a visual cosmetic defect on the top surface.The operator notified the finish lead, who contacted process and quality engineering for further evaluation.The image attached shows a augment returned from expiry with foam particles transferred from the packaging components.On (b)(6) 2021 12 parts were discovered with defects (captured on pc-001042377).On (b)(6) 2022, an additional 12 parts were discovered with defects (captured on pc-(b)(4)).Later on (b)(6) 2022, an additional 29 parts were discovered with defects (captured on pc-(b)(4)).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # : (b)(4).Investigation summary : examination of the returned sigma augment confirmed inner foam packaging residue/debris was found transferred to the surface of the device.The debris was found in raynham during repacking/resterilization of expired sterile product returned from the field.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the product and lot code was provided, a manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|