• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. ARTICULAR SURFACE FIXED BEARING CR RIGHT 14 MM HEIGHT; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. ARTICULAR SURFACE FIXED BEARING CR RIGHT 14 MM HEIGHT; PROSTHESIS, KNEE Back to Search Results
Model Number 42-5210-005-14
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Loss of Range of Motion (2032)
Date of Event 04/23/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Concomitant medical product: femur cemented cruciate retaining (cr) standard right size 8, catalog#: 42502606402, lot#: 63947833.Tibia cemented 5 degree stemmed right size e, catalog#: 42532007102, lot#: 63947833.All poly patella cemented 32 mm diameter, catalog#: 42540000032, lot#: 64129910, optipac-s 60 refob bn cmt r, catalog#: 4711500396-1, lot#: a746d06425.Report source: foreign: (b)(6).Multiple mdr reports have been filed for this event.Please see associated report: 3007963827-2022-00057.Arthrofibrosis is defined as a complication of injury or trauma where an excessive scar tissue response leads to painful restriction of joint motion, with scar tissue forming within the joint and surrounding soft tissue spaces and persisting despite rehabilitation exercises and stretches.A manipulation of the joint is needed to loosen the scar tissue to regain range of motion.The necessary intervention is a non-invasive technique performed for treating stiffness and poor range of motion following total joint replacement.There is no product involvement and performed due to scar tissue only.
 
Event or Problem Description
It was reported via a clinical study that a study patient underwent a primary total knee arthroplasty.Approximately three months post-implantation the patient underwent a closed brisement procedure due to a loss in range of motion.Attempts have been made and no further information has been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE FIXED BEARING CR RIGHT 14 MM HEIGHT
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13593881
Report Number0001822565-2022-00595
Device Sequence Number11727618
Product Code JWH
UDI-Device Identifier00889024239739
UDI-Public(01)00889024239739(17)240831(10)63455026
Combination Product (Y/N)N
Initial Reporter CountryDA
PMA/510(K) Number
K113369
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2024
Device Model Number42-5210-005-14
Device Catalogue Number42521000514
Device Lot Number63455026
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/01/2022
Initial Report FDA Received Date02/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/06/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight88 KG
-
-