| Model Number |
9004051 |
| Medical Device Problem Codes |
Unable to Obtain Readings (1516); Calibration Problem (2890)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there is an unstable co2 reading and calibration.No additional information.
|
| |
|
Event or Problem Description
|
|
It was reported that there is an unstable co2 reading and calibration.No additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
A product sample was received for evaluation.Visual and functional testing were performed.Visual inspection found a broken barrel from the filter slot attachment and the tamper seal was broken.Functional testing found a defective c02 bench which caused the unstable c02 reading and calibration error.The root cause of the reported issue was found to be a malfunctioning co2 sensor from wear of 20 years usage.The monitor was last serviced over 6 years ago in jan 2016 and had passed all functional tests per ro# 656519.Further device history record (dhr) review is not relevant because the results of the complaint investigation do not indicate a problem with the initial manufacture or prior repair of the device.
|
| |
|