• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR PLUS 30 PRIORITY PACK¿ ACCESSORY KIT; ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR PLUS 30 PRIORITY PACK¿ ACCESSORY KIT; ACCESSORIES Back to Search Results
Model Number 1000185
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/03/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device will not be returned for evaluation; the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.Date of event estimated.
 
Event or Problem Description
It was reported based on a comment that during the procedure the indeflator was noted to be pulling air into the chamber.A leak seems to be coming from the tubing that connects to the chamber.The indeflator was changed to a new indeflator to complete the procedure.There was no adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
Product performance engineering reviewed the incident information; however, the product was not returned to abbott vascular for analysis.The investigation determined the reported difficulties appear to be a potential product quality issue.On feb 25th, 2022 abbott vascular decided to initiate a voluntary field action for this product.Abbott vascular submitted medwatch # 2024168-2022-01831 on march 4, 2022 with notification of the voluntary recall in h7, (remedial action initiated).Corrective action has been initiated per site operating procedures.Field safety corrective action is required for specific lots of 20/30 and plus 30 indeflators and associated priority packs: 20/30 priority pack with copilot, 20/30 priority pack, priority pack 20/30 w/115 rhv, ppak 20/30 with rhv, plus 30, ppakplus30, ppakplus30 w/115 rhv.The product will continue to be trended.This action is being taken due to an increase in the complaint trend for reported leak/splash and loose or intermittent connection.20/30 indeflators are at an increased risk of leaking due to a gap in the hose snap fitting.Stopcocks are at an increased risk of leaking due to a higher tendency for loose connections when not connected properly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLUS 30 PRIORITY PACK¿ ACCESSORY KIT
Common Device Name
ACCESSORIES
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13595088
Report Number2024168-2022-01989
Device Sequence Number5125561
Product Code MAV
UDI-Device Identifier08717648013607
UDI-Public08717648013607
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K962495
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 03/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2022
Device Model Number1000185
Device Catalogue Number1000185
Device Lot Number60315910
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/04/2022
Supplement Date Received by Manufacturer02/25/2022
Initial Report FDA Received Date02/24/2022
Supplement Report FDA Received Date03/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
UNK
Patient Sequence Number1
Patient SexUnknown
-
-