• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL OG CMPLX XTRASOFT COIL 4X6; NEUROVASCULAR EMBOLIZATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL OG CMPLX XTRASOFT COIL 4X6; NEUROVASCULAR EMBOLIZATION DEVICE Back to Search Results
Model Number 640CX0406
Medical Device Problem Code Stretched (1601)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/03/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported by a healthcare professional that during an endovascular coil embolization in a male patient, a 3mm x 4cm orbit galaxy complex xtrasoft coil (640cx0304/30615573) was found to be prematurely detached when the package was opened, prior to use.The 4mm x 6cm orbit galaxy complex xtrasoft coil (640cx0406/30562629) became stretched during deployment of the coil in the mca, which required use of a snare for removal.The procedure was successfully completed with no patient injury.The surgery was delayed 10-20 minutes due to the event.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Procode: krd/hcg.(b)(6).The device is available to be returned for evaluation and testing.However, it has not been received to date as indicated as ¿other¿ in this section as the reason for non-evaluation.If the device returns, a device investigation will be performed.This information was not provided in the reported event or available at the time of report submission.The company is seeking this information through the event investigation.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.This is one of two products involved with the complaint and the associated manufacturer report number is 3008114965-2022-00145.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.Complaint conclusion: it was reported by a healthcare professional that during an endovascular coil embolization in a male patient, a 3mm x 4cm orbit galaxy complex xtrasoft coil (640cx0304/30615573) was found to be prematurely detached when the package was opened, prior to use.The 4mm x 6cm orbit galaxy complex xtrasoft coil (640cx0406/30562629) became stretched during deployment of the coil in the mca, which required use of a snare for removal.The procedure was successfully completed with no patient injury.The surgery was delayed 10-20 minutes due to the event.A non-sterile og cmplx xtra soft coil 3x4 was received inside of a pouch.The embolic coil component was outside of the introducer and severely stretched, the blue stretch resistant fiber was snapped off.No other damages could be noted on the rest of the components.A manufacturing record evaluation was performed for the finished device [30562629] number, and no non-conformance's related to the reported complaint condition were identified.The device was returned to cerenovus for further evaluation.Upon arrival, it was noted that the embolic coil component was outside of the introducer and severely stretched, the blue stretch resistant fiber was snapped off.No other damages could be noted on the rest of the components.Coil stretching is a known potential issue associated with the use of this device.The instructions for use (ifu) provides proper handling instructions for the device to prevent such issue from occurring.The customer complaint was confirmed based on the finding noted on the embolic coil component, however, no contributing factors could be identified from the device.Since the concomitant microcatheter (prowler select plus) was not facilitated in the lab, no further evaluation could be conducted.A manufacturing record evaluation was performed, and no non-conformance was found during the review.It should be noted that product failure could be caused by multiple factors.The instructions for use do contain the following precaution: ¿ never advance, withdraw, or torque the delivery tube against resistance without first determining the cause of resistance under fluoroscopy.Manipulation of the delivery tube against resistance can cause damage and/or premature detachment of the coil.Coil unraveling/stretching is classified as an mdr reportable malfunction as this failure could cause coil premature detachment, separation, herniation into parent vessel, with the potential for non-target site embolization, ischemia or infarct.In this case, the event required use of a snare to preclude patient complications, such as non-target site embolization, ischemia, or infarct.The file will be re-reviewed if additional information is received at a later date.As part of the cerenovus quality process, all devices are manufactured, inspected, and released to approved specifications.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OG CMPLX XTRASOFT COIL 4X6
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
Manufacturer (Section G)
CODMAN & SHURTLEFF, INC. (JUAREZ)
circuito interior norte #1820
parque industrial salvarcar
juarez chihuahua 32574
MX   32574
Manufacturer Contact
gabriel alfageme
31 technology dr
irvine, CA 92618
9497898687
MDR Report Key13595137
Report Number3008114965-2022-00147
Device Sequence Number20297719
Product Code KRD
UDI-Device Identifier10886704030621
UDI-Public10886704030621
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K093973
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2023
Device Model Number640CX0406
Device Catalogue Number640CX0406
Device Lot Number30562629
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/03/2022
Supplement Date Received by Manufacturer04/11/2022
Initial Report FDA Received Date02/24/2022
Supplement Report FDA Received Date04/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OG CMPLX XTRASOFT COIL 4X6; UNKPROWLER
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
Patient Weight75 KG
-
-