Model Number CD2311-36Q
Medical Device Problem Code Pacing Asynchronously (1441)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/15/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Related manufacturer reference numbers: 2017865-2022-04019.It was reported that the patient presented in clinic for follow-up.Upon interrogation, it was found that the implantable cardioverter defibrillator (icd) exhibited competitive atrial pacing and the atrial lead exhibited under-sensing as well as inappropriate automatic mode switch due to noise over-sensing.The icd and the atrial lead were both explanted and replaced.There were no patient consequences.
 
Additional Manufacturer Narrative
The reported event of pacing anomaly could not be confirmed in the lab.Interrogation of the device revealed the device was at the elective replacement indicator when received.A longevity calculation was performed and found the battery depletion was normal based on the device usage.The device was tested on the bench and using automated testing equipment, and no anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELLIPSE DR ICD
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key13595620
Report Number2017865-2022-04018
Device Sequence Number1931070
Product Code NVZ
UDI-Device Identifier05414734505994
UDI-Public05414734505994
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P910023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2014
Device Model NumberCD2311-36Q
Device Catalogue NumberCD2311-36Q
Device Lot Number4038935
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/17/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/15/2022
Supplement Date Received by Manufacturer03/11/2022
Initial Report FDA Received Date02/24/2022
Supplement Report FDA Received Date03/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/16/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DURATA.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient SexMale
Patient Weight139 KG
-
-