Model Number N/A
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/22/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported "pg feel dull and won't puncture skin." (b)(6) 2022-three devices were returned for evaluation.Each device exhibited rough/split transitions between the catheter tips and needle shafts.This report addresses the third device.
 
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of difficult powerglide catheter insertion was confirmed; however, the root cause was not identified.The product returned for evaluation was three 18ga x 10cm powerglide pro midline catheter assemblies.The product returned for evaluation was three 18ga x 10cm powerglide pro midline catheter assemblies.The first sample (sample 1) exhibited use residues throughout.The catheter was advanced and the safety mechanism was engaged.The catheter shaft exhibited curved shape memory.A split was observed in the catheter shaft near the distal tip.Microscopic inspection of sample 1 revealed deformation of the distal tip.Inspection of the split revealed a partially glossy fracture surface consistent with perforation by the introducer needle tip.The second sample (sample 2) was received fully assembled.Blood residue was observed within the catheter shaft.The third sample (sample 3) was received fully assembled and appeared free of obvious use residues.Microscopic inspection of samples 2 and 3 revealed irregular catheter tips.The transitions between the catheter tips and needle shafts were rough.The rough transitions would likely result in resistance when attempting to advance the catheters.The rough transitions appeared to be caused either by irregular catheter tip formation during the catheter tipping process or subsequent tip damage.Photographs have been forwarded to the manufacturing site for further evaluation.A lot history review (lhr) of refw1235 showed three other similar product complaint(s) from this lot number.
 
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of difficult powerglide catheter insertion was confirmed; however, the root cause was not identified.The product returned for evaluation was three 18ga x 10cm powerglide pro midline catheter assemblies.The product returned for evaluation was three 18ga x 10cm powerglide pro midline catheter assemblies.The first sample (sample 1) exhibited use residues throughout.The catheter was advanced and the safety mechanism was engaged.The catheter shaft exhibited curved shape memory.A split was observed in the catheter shaft near the distal tip.Microscopic inspection of sample 1 revealed deformation of the distal tip.Inspection of the split revealed a partially glossy fracture surface consistent with perforation by the introducer needle tip.The second sample (sample 2) was received fully assembled.Blood residue was observed within the catheter shaft.The third sample (sample 3) was received fully assembled and appeared free of obvious use residues.Microscopic inspection of samples 2 and 3 revealed irregular catheter tips.The transitions between the catheter tips and needle shafts were rough.The rough transitions would likely result in resistance when attempting to advance the catheters.The rough transitions appeared to be caused either by irregular catheter tip formation during the catheter tipping process or subsequent tip damage.Photographs have been forwarded to the manufacturing site for further evaluation.A lot history review (lhr) of refw1235 showed three other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported "pg feel dull and won't puncture skin." (02/23/2022-three devices were returned for evaluation.Each device exhibited rough/split transitions between the catheter tips and needle shafts.This report addresses the third device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PG PRO 18G 10CM BASIC
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kelsey erickson
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key13595698
Report Number3006260740-2022-00445
Device Sequence Number12226770
Product Code FOZ
UDI-Device Identifier00801741110511
UDI-Public(01)00801741110511
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K153280
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberF118100
Device Lot NumberREFW1235
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/23/2022
Initial Report FDA Received Date02/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-