MAUDE Adverse Event Report: MICROVENTION, INC. AZUR CX 18D-LONGER; VASCULAR EMBOLIZATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION, INC. AZUR CX 18D-LONGER; VASCULAR EMBOLIZATION DEVICE Back to Search Results
Model Number 45-780308-L
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/27/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during embolization treatment of a left internal mammary artery collateral, the coil implant became detached unexpectedly in the collateral.There was no patient injury or intervention.It was reported that the patient is stable.
 
Additional Manufacturer Narrative
A search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was not returned to the manufacturer for analysis.The alleged product issue cannot be confirmed.If the device is received at a later date, an investigation will be performed and a supplemental report will be submitted.The instructions for use (ifu) identifies premature coil detachment as a potential complication associated with use of the device.
 
Additional Manufacturer Narrative
The pusher was the only component received for evaluation.No indication using a detachment controller was found on the pusher's heater coil.The investigation found the pusher's monofilament with a tensile break shape at the tip which is consistent with the device experiencing excessive force that exceeded the strength of the monofilament causing the implant to separate from the pusher.The physical evaluation of the device could not identify the conditions or circumstances that led to the damage, but the damage is consistent with the device experiencing forces over specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZUR CX 18D-LONGER
Common Device Name
VASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise drive
aliso viejo, CA 92656
7142478000
MDR Report Key13595788
Report Number2032493-2022-00073
Device Sequence Number1857148
Product Code KRD
UDI-Device Identifier00816777024398
UDI-Public(01)00816777024398(11)210412(17)260331(10)2104125Z8
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
K123384
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2026
Device Model Number45-780308-L
Device Lot Number2104125Z8
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/28/2022
Supplement Date Received by Manufacturer03/30/2022
Initial Report FDA Received Date02/24/2022
Supplement Report FDA Received Date04/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-