MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD MAXZERO MULTI-FUSE EXTENSION SET WITH NEEDLELESS CONNECTOR(S); INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD MAXZERO MULTI-FUSE EXTENSION SET WITH NEEDLELESS CONNECTOR(S); INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number MZ9273
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the bd maxzero multi-fuse extension set with needleless connector(s) experienced leakage.The following information was provided by the initial reporter: leaking of the filter.
 
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d9: device available for eval yes, d9: returned to manufacturer on: 01-mar-2022.H6: investigation summary the customer reported leaking from the filter, and returned one used sample of material mz9273.The set leaked from the filter air vent, and the complaint is verified.Further investigation was done by shipping the sample to the supplier of the filter.The supplier was not able to attribute a root cause to the vent leak.A device history record review for model mz9273 lot number 21116090 was performed.The search showed that a total of (b)(4) units in 1 lot number was built on 18nov2021.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.H3 other text : see h10.
 
Event or Problem Description
It was reported that the bd maxzero multi-fuse extension set with needleless connector(s) experienced leakage.The following information was provided by the initial reporter: leaking of the filter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MAXZERO MULTI-FUSE EXTENSION SET WITH NEEDLELESS CONNECTOR(S)
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13595823
Report Number9616066-2022-00175
Device Sequence Number1956792
Product Code FPA
UDI-Device Identifier50885403240858
UDI-Public50885403240858
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K140831
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date11/18/2024
Device Model NumberMZ9273
Device Catalogue NumberMZ9273
Device Lot Number21116090
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/15/2022
Supplement Date Received by Manufacturer04/14/2022
Initial Report FDA Received Date02/24/2022
Supplement Report FDA Received Date04/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-