THE SPECTRANETICS CORPORATION SPECTRANETICS TIGHTRAIL ROTATING DILATOR SHEATH; DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION Back to Search Results
Model Number 545-513
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient''s weight unk.Relevant tests/laboratory data unk.Device lot number, expiration date unk.The device was discarded, thus no investigation could be completed.Device manufacture date unk because lot number unk.
 
Event or Problem Description
A lead extraction procedure commenced to remove a right atrial (ra) and a right ventricular (rv) lead due to infection and rv lead fracture (implanted in (b)(6)).The ra lead was successfully removed.The rv lead was severely calcified.The physician first tried a spectranetics 14f glidelight laser sheath and then upsized to a 16f glidelight device with use of a spectranetics visisheath dilator sheath.The physician switched to a spectranetics 13f tightrail rotating dilator sheath due to slow progress.The tightrail device got stuck on the lead''s coil near the brachiocephalic region and the physician was unable to back it off the lead.It was noted that the tightrail device became stuck at a point on the rv lead where it had previously been fractured.The physician decided to just pull the lead with the tightrail on it.Pulling with traction showed no hemodynamic changes so he continued to pull.The rv lead eventually freed and the procedure was completed with no patient harm.This report captures the tightrail device which became entrapped on the rv lead, potential for harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRANETICS TIGHTRAIL ROTATING DILATOR SHEATH
Common Device Name
DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION
Manufacturer (Section D)
THE SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer (Section G)
THE SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer Contact
barbara creel
9965 federal drive
colorado springs, CO 80921
MDR Report Key13595884
Report Number1721279-2022-00032
Device Sequence Number1930480
Product Code DRE
UDI-Device Identifier00813132020897
UDI-Public00813132020897
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K150360
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number545-513
Device Catalogue Number545-513
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/28/2022
Initial Report FDA Received Date02/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BOSTON SCIENTIFIC 1500 RV ICD LEAD; SPECTRANETICS 14F GLIDELIGHT LASER SHEATH; SPECTRANETICS 16F GLIDELIGHT LASER SHEATH; SPECTRANETICS CVX-300 EXCIMER LASER SYSTEM; SPECTRANETICS LEAD LOCKING DEVICES; SPECTRANETICS VISISHEATH DILATOR SHEATH; ST. JUDE MEDICAL 1688TC RA PACING LEAD
Patient Age35 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-