• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPICC 5F DUAL-LUMEN POLYURETHANE CATHETER RADIOLOGY BASIC TRAY; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPICC 5F DUAL-LUMEN POLYURETHANE CATHETER RADIOLOGY BASIC TRAY; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/28/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of refw0171 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported 5f dl radiology picc leaked at the casing on the purple hub.The device was not used on a patient.
 
Additional Manufacturer Narrative
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of refw0171 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported 5f dl radiology picc leaked at the casing on the purple hub.The device was not used on a patient.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of leakage at the purple luer adapter was unconfirmed because the problem could not be reproduced.The product returned for evaluation was one 5fr d/l powerpicc catheter.The sample appeared free of obvious use residues.The catheter was untrimmed.Microscopic inspection of the sample was unremarkable.Both luer adapters appeared well-formed and unremarkable.Inspection of both luer adapters using an iso taper gauge revealed both luers to be within manufacturing specifications.An attempt to infuse water through the sample using a 12ml syringe revealed both lumens to be patent to infusion and aspiration with no observed leaks.No leaks were observed during sustained (>15 seconds) hydraulic pressurization.No deficiencies were discovered during evaluation of the returned sample.Consequently this complaint is unconfirmed at this time.H3 other text : evaluation findings are in section h.11.
 
Event or Problem Description
It was reported 5f dl radiology picc leaked at the casing on the purple hub.The device was not used on a patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC 5F DUAL-LUMEN POLYURETHANE CATHETER RADIOLOGY BASIC TRAY
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kelsey erickson
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key13596013
Report Number3006260740-2022-00448
Device Sequence Number20295672
Product Code LJS
UDI-Device Identifier00801741027963
UDI-Public(01)00801741027963
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K051672
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number3275355
Device Lot NumberREFW0171
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/04/2022
Supplement Date Received by Manufacturer03/30/2022
Initial Report FDA Received Date02/24/2022
Supplement Report FDA Received Date04/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-