• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® URETHRAL CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® URETHRAL CATHETER Back to Search Results
Model Number 120616
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/29/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that few of intermittent catheters was found discolored inside the packaging.
 
Additional Manufacturer Narrative
The reported event was confirmed however the cause was unknown.One sample was confirmed to exhibit the reported failure.As the device had signs of degradation, the reported event is confirmed to have a relationship with the device.As the cause of this issue was unknown, it could not be determined if the device met specifications.As the product was not used on a patient, it was not used for diagnostic or treatment purposes.An orange coude tip catheter was returned unopened in original packaging.The catheter was noted to have multiple discoloration marks.Catheter discoloration was confirmed to be caused by oxidation.As is it unable to determine when the oxidation occurred the reported event will be confirmed cause unknown.Photo samples of the catheter was also returned.The samples also showed discoloration of the catheter caused by oxidation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "storage: store catheters at room temperature away from direct exposure to light, preferably in the original box".
 
Additional Manufacturer Narrative
The reported event was confirmed however the cause was unknown.One sample was confirmed to exhibit the reported failure.As the device had signs of degradation, the reported event is confirmed to have a relationship with the device.As the cause of this issue was unknown, it could not be determined if the device met specifications.As the product was not used on a patient, it was not used for diagnostic or treatment purposes.An orange coude tip catheter was returned unopened in original packaging.The catheter was noted to have multiple discoloration marks.Catheter discoloration was confirmed to be caused by oxidation.As is it unable to determine when the oxidation occurred the reported event will be confirmed cause unknown.Photo samples of the catheter was also returned.The samples also showed discoloration of the catheter caused by oxidation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "storage: store catheters at room temperature away from direct exposure to light, preferably in the original box".
 
Event or Problem Description
It was reported that few of intermittent catheters was found discolored inside the packaging.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® URETHRAL CATHETER
Common Device Name
URETHRAL CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key13596169
Report Number1018233-2022-00858
Device Sequence Number1930492
Product Code KOD
UDI-Device Identifier00801741029974
UDI-Public(01)00801741029974
Combination Product (Y/N)N
Initial Reporter StateND
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2024
Device Model Number120616
Device Catalogue Number120616
Device Lot NumberNGEQ1889
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/29/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/01/2022
Initial Report FDA Received Date02/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/23/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-