• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE; ELECTRODE, ELECTROSURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE; ELECTRODE, ELECTROSURGICAL Back to Search Results
Model Number WA47506S
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/24/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is ongoing.A supplemental report will be submitted upon receiving additional information.
 
Event or Problem Description
Olympus was informed of the following event from (b)(6) health services ((b)(6) mdip#(b)(4)): during a therapeutic fibroid resection, the tip of the olympus electrode loop broke inside the patient.The tip of the device was retrieved from the patient with an unspecified surgical instrument.The intended surgery was completed with a similar device.There was an unspecified surgical delay to remove the fragment and get a new device.There was no patient harm.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer¿s investigation and the results of the device history records (dhr) review.The dhr for this device were reviewed and all records indicated the product was manufactured according to all applicable procedures and met final product release criteria.No abnormalities were found.A definitive root cause was not identified.Based on the available information, the legal manufacturer determined the probable cause of the failure is likely due to wear and tear; he loop at the distal end of the electrode wears out during use and may break, burn or melt.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE
Common Device Name
ELECTRODE, ELECTROSURGICAL
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer Contact
daniel wladow
kuehnstrasse 61
hamburg 22045
GM   22045
4940669662955
MDR Report Key13596314
Report Number9610773-2022-00080
Device Sequence Number12220939
Product Code JOS
UDI-Device Identifier14042761085042
UDI-Public14042761085042
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K161600
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup
Report Date (Section B) 04/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/14/2025
Device Model NumberWA47506S
Device Catalogue NumberWA47506S
Device Lot Number1000067135
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/09/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/25/2022
Supplement Date Received by Manufacturer03/14/2022
Initial Report FDA Received Date02/24/2022
Supplement Report FDA Received Date04/11/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/09/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNSPECIFIED SURGICAL INSTRUMENT
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-