| Model Number |
SD980.014 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Difficulty Chewing (4522); Implant Pain (4561); Insufficient Information (4580)
|
| Date of Event |
11/04/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Maxilla plate broke in two locations and will be removed.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation ongoing.Device evaluated by mfr: device still in patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation ongoing.Device evaluated by mfr: device still in patient.
|
| |
|
Event or Problem Description
|
|
Maxilla plate broke in two locations and will be removed.
|
| |
|
Additional Manufacturer Narrative
|
|
Patient condition required big advancement of the maxilla and he was grinding his teeth at night, placing additional forces on the plate.H3: plate cannot be located at the hospital.
|
| |
|
Search Alerts/Recalls
|