• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL ADVISOR¿ FL, CIRCULAR MAPPING CATHETER, SENSOR ENABLED¿; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL ADVISOR¿ FL, CIRCULAR MAPPING CATHETER, SENSOR ENABLED¿; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number D-AVSE-DF10-F15
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Cardiac Perforation (2513)
Date of Event 02/14/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.Per the ifu, vascular perforation is an inherent risk of any electrode placement.  .
 
Event or Problem Description
During an atrial fibrillation ablation procedure, a cardiac tamponade was diagnosed via echocardiogram.Initially when starting to map the left atrium of the patient, the patient reported pain and moved.The mapping catheter was on the posterior wall of the left atrium between the two pulmonary veins, specifically closer to the inferior pulmonary vein, when the pain was reported.An echocardiogram was then performed which diagnosed a pericardial effusion.The effusion was drained and medication was administered to reverse the heparin the patient initially received.Then patient was then stabilized and the procedure was not continued.There were no patient symptoms.There were no performance issues with the abbott device.
 
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.Per the ifu, vascular perforation is an inherent risk of any electrode placement.  .
 
Event or Problem Description
During an atrial fibrillation ablation procedure, a cardiac tamponade was diagnosed via echocardiogram.Initially when starting to map the left atrium of the patient, the patient reported pain and moved.The mapping catheter was on the posterior wall of the left atrium between the two pulmonary veins, specifically closer to the inferior pulmonary vein, when the pain was reported.An echocardiogram was then performed which diagnosed a pericardial effusion.The effusion was drained and medication was administered to reverse the heparin the patient initially received.Then patient was then stabilized and the procedure was not continued.There were no patient symptoms.There were no performance issues with the abbott device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVISOR¿ FL, CIRCULAR MAPPING CATHETER, SENSOR ENABLED¿
Common Device Name
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key13598866
Report Number3008452825-2022-00068
Device Sequence Number14917110
Product Code DRF
UDI-Device Identifier05415067025975
UDI-Public05415067025975
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K160335
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2023
Device Model NumberD-AVSE-DF10-F15
Device Catalogue NumberD-AVSE-DF10-F15
Device Lot Number8146625
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/14/2022
Initial Report FDA Received Date02/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight53 KG
-
-