Model Number M00555540
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that an acquire needle was used in the pancreas during an endoscopic ultrasound (eus) procedure performed on an unknown date.During the procedure, the distal tip of the needle detached inside the patient and was not removed.Reportedly, the physician followed up with the patient and the patient is asymptomatic.The procedure was completed with another acquire needle.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
The exact date of the event is unknown.The provided event date was chosen as a best estimate based on the date that the manufacturer became aware of the event.(b)(4).Investigation results: the returned acquire needle was analyzed, a visual evaluation noted the working length was kinked and the tip of the needle was broken.The device came without the stylet.No other issues were noted.The reported event was confirmed.Based on all available information and the condition of the returned device, it is most likely that the manipulation or the technique used by the physician in conjunction with the patient anatomy and excess force applied during the procedure could have affected the device operation, resulting in the break on the needle.The investigation concluded the most probable root cause is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACQUIRE
Common Device Name
ENDOSCOPIC ULTRASOUND SYSTEM, GASTROENTEROLOGY-UROLOGY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13598877
Report Number3005099803-2022-00625
Device Sequence Number5857235
Product Code ODG
UDI-Device Identifier08714729931782
UDI-Public08714729931782
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K160845
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/19/2024
Device Model NumberM00555540
Device Catalogue Number5554
Device Lot Number0028238001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/28/2022
Initial Report FDA Received Date02/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-