| Model Number |
R5100052 |
| Medical Device Problem Code |
Pumping Problem (3016)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/26/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product analysis: the reported issue of pump 1 was intermittently not working was verified during field service.During service the technician noted that when turning on the instrument the motor would spin during the self-test.The technician pressed the start button for the pm4 which did not work.The technician then swapped the connectors on the sin-ui board and they were able to turn the motor on with a pm2's start button while pm2 itself would not start with pm4's start button.The issue was resolved by replacing the overlay for pm4 (pn r5903049 ).Post-repair testing was performed per specifications.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that during use of this performer instrument, the customer reported that pump 1 was intermittently not working.The instrument primed the cardioplegia pump and delivered the first rx correctly.The second delivery stopped at 100ml.The customer then had to use the hand-crank before restarting the instrument with no issues.The instrument was replaced with a back up and there was no reported adverse patient effect.Additional information was requested to confirm how long the handcrank was in use for, however this is unavailable.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: the reported issue of pump 1 was intermittently not working was verified during field service.During service the technician noted that when turning on the instrument the motor would spin during the self-test.The technician pressed the start button for the pm4 which did not work.The technician then swapped the connectors on the sin-ui board and they were able to turn the motor on with a pm2's start button while pm2 itself would not start with pm4's start button.The issue was resolved by replacing the overlay for pm4 (pn r5903049).Post-repair testing was performed per specifications.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that during use of this performer instrument, the customer reported that pump 1 was intermittently not working.The instrument primed the cardioplegia pump and delivered the first rx correctly.The second delivery stopped at 100ml.The customer then had to use the hand-crank before restarting the instrument with no issues.The instrument was replaced with a back up and there was no reported adverse patient effect.Additional information was requested to confirm how long the handcrank was in use for, however this is unavailable.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|