| Model Number |
402810 |
| Medical Device Problem Code |
Signal Artifact/Noise (1036)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
During an atrioventricular nodal reentrant tachycardia procedure, due to a noise issue and the troubleshooting involved, a prolonged procedure occurred.Initially the cable and generator were exchanged, but the issue remained.The catheter was then exchanged and the issue was resolved.There were no adverse consequences to the patient.
|
| |
|