Model Number N/A
Medical Device Problem Code Activation Problem (4042)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of refu1390 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported "needle would not safety." additional info.Received (b)(6) 2022: accucath placed without difficulty, when button depressed to safety needle, it would not retract.Button was fully depressed.No patient harm.
 
Additional Manufacturer Narrative
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of refu1390 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported "needle would not safety." additional info.Received (b)(6) 2022: accucath placed without difficulty, when button depressed to safety needle, it would not retract.Button was fully depressed.No patient harm.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of safety mechanism failure was inconclusive because of the nature of the submitted evidence.The product returned for evaluation was two photographs which depicted a 20ga x 2.25¿ accucath ace peripheral iv catheter assembly.The catheter had been advanced and was not visible in the photographs.In both photographs, the guidewire appeared to be fully advanced.The guidewire exhibited curved shape memory.The needle was not withdrawn into the housing.No obvious device deficiency was observed during examination of the returned photographs.The needle was not withdrawn; however, it could not be determined if the safety button was pressed, nor could the safety mechanism be thoroughly inspected or tested.Consequently this complaint is inconclusive at this time.H3 other text : evaluation findings are in section h.11.
 
Event or Problem Description
It was reported "needle would not safety." additional info.Received: accucath placed without difficulty, when button depressed to safety needle, it would not retract.Button was fully depressed.No patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCUCATH INTRAVASCULAR CATHETER 20GX2.25
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kelsey erickson
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key13599088
Report Number3006260740-2022-00463
Device Sequence Number14886148
Product Code FOZ
UDI-Device Identifier00801741111006
UDI-Public(01)00801741111006
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K153298
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberAC0202252
Device Lot NumberREFU1390
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/08/2022
Supplement Date Received by Manufacturer03/30/2022
Initial Report FDA Received Date02/25/2022
Supplement Report FDA Received Date04/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-