| Model Number |
24690 |
| Medical Device Problem Codes |
Material Rupture (1546); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon pinhole occurred.The 95% stenosed, 6mm target lesion was located in the moderately tortuous and severely calcified superficial femoral artery.Pre-inflation was performed with a 3mm balloon.After the non-boston scientific guidewire was crossed to the lesion, a 5.0mmx220mmx150cm (4f) sterling balloon catheter was advanced for dilatation.However, during the first inflation at 10 atmospheres for 30 seconds, the balloon twisted.The device was removed.In an attempt to fix the twisted balloon outside of the patient, the guidewire was advanced in and out of the balloon resulting in the guidewire penetrating through the balloon.The procedure was completed with a different device.There were no patient complications nor injuries reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Initial reporter city: (b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
Initial reporter city: (b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that balloon pinhole occurred.The 95% stenosed, 6mm target lesion was located in the moderately tortuous and severely calcified superficial femoral artery.Pre-inflation was performed with a 3mm balloon.After the non-boston scientific guidewire was crossed to the lesion, a 5.0mmx220mmx150cm (4f) sterling balloon catheter was advanced for dilatation.However, during the first inflation at 10 atmospheres for 30 seconds, the balloon twisted.The device was removed.In an attempt to fix the twisted balloon outside of the patient, the guidewire was advanced in and out of the balloon resulting in the guidewire penetrating through the balloon.The procedure was completed with a different device.There were no patient complications nor injuries reported.
|
| |
|
Additional Manufacturer Narrative
|
|
E1- initial reporter city: (b)(6).Device evaluated by mfr.: returned product consisted of a sterling balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed no damages.Microscopic examination revealed a hole in the shaft 10.2cm from the hub.The device was functionally tested by inflating it but the device was leaking from the hole so the balloon could not be verified to have any ruptures.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon pinhole occurred.The 95% stenosed, 6mm target lesion was located in the moderately tortuous and severely calcified superficial femoral artery.Pre-inflation was performed with a 3mm balloon.After the non-boston scientific guidewire was crossed to the lesion, a 5.0mmx220mmx150cm (4f) sterling balloon catheter was advanced for dilatation.However, during the first inflation at 10 atmospheres for 30 seconds, the balloon twisted.The device was removed.In an attempt to fix the twisted balloon outside of the patient, the guidewire was advanced in and out of the balloon resulting in the guidewire penetrating through the balloon.The procedure was completed with a different device.There were no patient complications nor injuries reported.
|
| |
|