• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER MANUFACTURING B.V. FEMORAL HEAD STERILE PRODUCT DO NOT RESTERILIZE 12/14 TAPER; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER MANUFACTURING B.V. FEMORAL HEAD STERILE PRODUCT DO NOT RESTERILIZE 12/14 TAPER; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Bacterial Infection (1735); Hematoma (1884); Impaired Healing (2378)
Date of Event 08/07/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation as the product remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Concomitant products: part # unknown / unk tm 64mm cup/ lot # unknown ; part #unknown / unk liner/ lot # unknown ; part #unknown / unk screws/ lot # unknown/ quantity 7; part# 01.00102.220 / wagner sl revision stem 20/ / lot # 2940737.Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2022-00226, 0001822565-2022-00229, 0001822565-2022-00230, 0001822565-2022-00269, 0001822565-2022-00270, 0001822565-2022-00271, 0001822565-2022-00273, 0001822565-2022-00274, 0001822565-2022-00275.
 
Event or Problem Description
It was reported that patient underwent total left hip arthroplasty approximately.Subsequently, the patient underwent two separate revisions of the acetabular and femoral components due to loosening, approximately 4 years later.Approximately three weeks after the second revision, the patient experienced a postop hematoma which was surgically debrided without complication.Intraop cultures grew staphylococcus epidermis indicating a superficial infection.The debridement did not extend into the joint, and no product was exchanged.After changing the postop antibiotics, the patient healed without further complications.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.D10: 00875706402 ¿ continuum shell ¿ unknown lot; 00875201636 ¿ xlpe liner ¿ unknown lot; 00625006520 ¿ bone screw ¿ unknown lot; 00625006525 ¿ bone screw ¿ unknown lot; 00625006535 ¿ bone screw ¿ unknown lot; 00625006535 ¿ bone screw ¿ unknown lot; 00625006530 ¿ bone screw ¿ unknown lot; 00625006520 ¿ bone screw ¿ unknown lot; 00625006540 ¿ bone screw ¿ unknown lot.It was reported that a patient experienced wound dehiscence and hematoma from that was debrided and repaired approximately three weeks post hip revision.The cultures from the procedure were positive for staph epidermis which is a common skin flora.It is expected that a wound heals in stages and should be of normal appearance related to the timeframe since the incision was made.A surgical wound should be well approximated without redness, warmth, swelling and/or purulent drainage for the duration of its healing.The expression wound concerns or non-healing wound would imply that the appearance of the wound deviates from what a surgical wound should appear as.It may be red, have drainage, additional pain, warmth and swelling as well as healing time may be delayed.This deviation signifies an alteration in the wound healing process.A hematoma is a mass of clotted blood that forms in a tissue, organ, or body space.A hematoma can be associated with pain, swelling, ecchymosis, serosanguinous or bloody drainage after a recent surgical procedure.The development of a postoperative hematoma after a procedure can be correlated with the surgical procedure and perioperative anticoagulation therapy to prevent dvt (prophylaxis).Patients may be at increased risk for developing hematomas if they have received fresh-frozen plasma, vitamin k, or hormonal therapy as well.Most hematomas resolve on their own, without surgical intervention, while some others do not.Larger hematomas may need to be surgically evacuated in order to resolve.As timeframes of onset differ due to individual contributing factors, a specific timeframe of expected occurrence cannot be established.The presence of a hematoma increases patient risk for infection and wound complications.At the time of positive cultures, the femoral head had been in place for approximately three weeks, and part and lot are known.Superficial infections are considered a procedure related complication as no device has been reported as revised.There are multiple factors that may contribute to an infection that are outside the control of zimmer biomet, such as external factors, i.E.Hospital/surgical environment, provider related risk factors, and/or patient comorbidities/risk factors.During the investigation process a review of the sterile certifications were reviewed and found to be conforming with no applicable deviations, further eliminating the implanted devices as a potential source for the reported infection.Devices were verified to have gone through acceptable sterilization process following iso/aami/astm & eu published guidelines.As device has not been indicated as revised and there are no indications of a product or process issues identified affecting implant safety or effectiveness, therefore implanted products are not identified as the source or contributing to the reported infection.All other subforms/implants had been in place for approximately 8 months and would not be considered as causing or contributing to the recent superficial infection.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional event information to report at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL HEAD STERILE PRODUCT DO NOT RESTERILIZE 12/14 TAPER
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer (Section G)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13599440
Report Number0002648920-2022-00046
Device Sequence Number17590641
Product Code LPH
UDI-Device Identifier00889024144750
UDI-Public(01)00889024144750(17)271130(10)63800434
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K953337
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00801803605
Device Lot Number63800434
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/31/2022
Supplement Date Received by Manufacturer04/22/2022
Initial Report FDA Received Date02/25/2022
Supplement Report FDA Received Date04/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/13/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10; SEE H10 NARRATIVE
Outcome Attributed to Adverse Event Hospitalization; Other;
Patient SexMale
Patient Weight114 KG
-
-