| Model Number |
NSLG2C45A |
| Medical Device Problem Codes |
Difficult to Open or Close (2921); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/03/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch #: v95r5r.A manufacturing record evaluation was performed for the finished device and the manufacturing criteria were met prior to the release of this lot/batch.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during a robotic prostatectomy the device jaws locked inside the patient.While attempting to remove it the jaws broke.No fragments were generated and no x-ray was required.There were no patient consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch #: v95r5r.A manufacturing record evaluation was performed for the finished device and the manufacturing criteria were met prior to the release of this lot/batch.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during a robotic prostatectomy the device jaws locked inside the patient.While attempting to remove it the jaws broke.No fragments were generated and no x-ray was required.There were no patient consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 3/23/2022.Investigation summary: the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the nslg2c45a device was received the lower jaw detached at the welding point.Not all piece of the jaws were returned with the device.The device was connected to the generator and it was recognized.Because the jaw was damaged not all functional testing could be performed with the generator.A probable cause of the damage to the jaw could be not allowing thick and fibrous tissues to denature prior to i-blade advancement.As described in the ifu: do not use excessive force on the closing handle to close the jaws; grasp only as much tissue as will fit between the jaws where the current will pass.Greater amounts of tissue require more closing handle force.Excessive force could damage the device.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|
| |
|