| Model Number |
406116 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported leak could not be conclusively determined.The ifu states damage to the valve assembly may occur if inner catheter is withdrawn rapidly.
|
| |
|
Event or Problem Description
|
|
During the procedure, a fluid leak was noted from the hemostasis valve of the introducer when the guidewire was advanced through the valve.The procedure continued without exchanging the sheath with no consequences to the patient.
|
| |
|
Search Alerts/Recalls
|