| Model Number |
HEMI-PROSTHESIS ADAPTOR WITH SCREW |
| Medical Device Problem Code |
Defective Component (2292)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/29/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
On (b)(6) 2020 stryker trauma and extremities acquired tornier sas.An effort was made to align the acquired entity complaint files with stryker's policies and procedures.Stryker has performed retrospective review of complaints and the reportability decision was revised to report this incident under the mdr process.A device inspection shows that the union screw didn't fit well with an aequalis reversed fracture stem, the threading is too big.A review of the device history shows that the instruction of the medical devices evolution mention that the screw (dwb923 batch # 2623ag) must be scrapped.In this case the screw (dwb923 batch # 2623ag) has not been scrapped.The root cause can be attributed to a manufacturing defect.
|
| |
|
Event or Problem Description
|
|
During the surgical intervention of the fracture shoulder hemiprothesis system, the doctor requests a hemi prosthesis adapter ((b)(4)).It has two pieces: a screw and a plataform which fit to the humeral stem.The procedure was made follow the instructions and guides given for its correct implantation, but when the doctor tries to adjust the screw to the humeral stem it does not adjust properly, because the diameter of the screw is greater than the stem.
|
| |
|