On november 10st, 2020 stryker trauma and extremities acquired tornier sas.An effort was made to align the acquired entity complaint files with stryker's policies and procedures.Stryker has performed retrospective review of complaints and the reportability decision was revised to report this incident under the mdr process.The drilled tip of the drill bit was broken at approximately 38 mm from the tip.The drill tip is bent.Rupture of the shaft, between the drill bit sleeve and the drill bit tool, at the shoulder.To the touch cutting edges seem rather still sharp and tip is pointed.A review of the device history for the reported lot did not indicate any abnormalities.Based on data available, the "head part breakage was made intentionally by the surgeon when trying to pull out the stuck drill during the initial surgery".Then, the drill bit tip should have broken while pulling out the rest of the drill bit from the bone : excessive bending load should have finished breaking the drill bit in two pieces.Patient condition (very hard bone) could also be a reason why the drill bit got stuck into the bone.Moreover, according to the surgeon, this case was caused by a technical error.Furthermore, the broken part should have been removed during the initial surgery of the (b)(6) 2019.It is a surgeon' choice to have explanted it 6 days later.A device misuse (abnormal use) seems to be the most likely.
|
|
This patient had a surgery of aequalis reversed.When the ''dia 3mm drill bit (220mm)''was used, the drill tip part got stuck and couldn't pull out.It was managed to pull out forcedly, but the drill tip part got broked.This drill tip part couldn't be pulled out , so it had to remain inside the body.According to the surgeon,this case was caused by a technical error.He could not pull out because the drill tip part was clogged in the coracoid process of scapula.The drill is usually used with caution, but this time it took too long handling the drill.The revision surgery was carried out on june 6,and the broken drill tip was removed without any problem.
|