MAUDE Adverse Event Report: TORNIER S.A.S AEQUALIS ASCEND FLEX; PROSTHESIS SHOULDER POLYMER / METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER S.A.S AEQUALIS ASCEND FLEX; PROSTHESIS SHOULDER POLYMER / METAL Back to Search Results
Model Number STANDARD PTC STEM SIZE 7
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 06/21/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
On (b)(6) 2020 stryker trauma and extremities acquired tornier sas.An effort was made to align the acquired entity complaint files with stryker's policies and procedures.Stryker has performed retrospective review of complaints and the reportability decision was revised to report this incident under the mdr process.Additional devices listed in this report: cat: dwf520; aequalis ascend flex reversed tray th +0 offset 3.5mm; serial # (b)(4).Cat: dwf421b; aequalis ascend flex reversed insert dia.42 +6 12.5°; serial # (b)(4).A device inspection was not possible since the affected device was not returned and no other evidences were provided for investigation.A review of the device history for the reported lot did not indicate any abnormalities.After checking the sterile md release sheets of the complaint parts, no issue has been reported.All concerned results were compliant.No other complaint has been reported since today regarding infections with parts of the same lots implanted and regarding the traceability history of the lots.Infection can be related to several reasons: a previous surgery, a sub-acromial dead space, a poor sterile technique, more fragile patients or younger male patients who have a greater risk.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.
 
Event or Problem Description
Revision surgery for suspicion of infection not yet confirmed.The surgeon removed everything except baseplate (afraid to break everything because excellent bite).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AEQUALIS ASCEND FLEX
Common Device Name
PROSTHESIS SHOULDER POLYMER / METAL
Manufacturer (Section D)
TORNIER S.A.S
161 rue lavoisier
montbonnot saint martin, 38330
FR  38330
Manufacturer Contact
laurent ray
161 rue lavoisier
montbonnot saint martin 38330
FR   38330
MDR Report Key13606993
Report Number3000931034-2022-00062
Device Sequence Number19315378
Product Code KWS
UDI-Device Identifier03700386944666
UDI-Public03700386944666
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K122698
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2016
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Study
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date07/31/2018
Device Model NumberSTANDARD PTC STEM SIZE 7
Device Catalogue NumberDWF607B
Device Lot Number9217AM
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/21/2016
Initial Report FDA Received Date02/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/31/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age82 YR
Patient SexMale
Patient Weight75 KG
-
-