CARDINAL HEALTH N-C CONN TUBE 6MM X3.1M; TUBING, NONINVASIVE Back to Search Results
Model Number 8888301614
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/16/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event or Problem Description
The customer reported that single use suction tubing fell apart during surgical procedure.The blue rubber connector end of the suction tubing came apart from the clear tube during the procedure at a very critical time during when suction was needed urgently because of bleeding at the surgical site.The tubing was immediately removed from the surgical site and the anaesthetic technician opened another tubing straight away.The customer had safety concerns as the event happened at a critical time during the procedure.There was no patient harm reported during this incident.
 
Additional Manufacturer Narrative
The device history record was reviewed and indicated that the product was released accomplishing all quality standards.The device was manufactured on june 14, 2021.There were no physical samples or photos provided for evaluation.Because a sample was not returned, we were unable to perform a follow up investigation to include functional and visual evaluations to confirm the reported issue or determine the root cause.If samples are received at a later date, the complaint will be reopened, and the investigation will be updated accordingly.A corrective action is not applicable at this time.Functional testing and visual inspections are being performed according to our current quality standards and inspection procedures.We will continue to monitor the process, complaints, and feedback for any adverse trends that require immediate attention.This complaint will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
N-C CONN TUBE 6MM X3.1M
Common Device Name
TUBING, NONINVASIVE
Manufacturer (Section D)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX  22500
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key13607174
Report Number9612030-2022-03180
Device Sequence Number13935612
Product Code GAZ
UDI-Device Identifier20884527019653
UDI-Public20884527019653
Combination Product (Y/N)N
Initial Reporter CountryNZ
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8888301614
Device Catalogue Number8888301614
Device Lot Number2115939964
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/16/2022
Supplement Date Received by Manufacturer02/16/2022
Initial Report FDA Received Date02/25/2022
Supplement Report FDA Received Date04/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-