| Model Number |
ATELLICA CH LACTATE_2 (LAC_2) |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/06/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Discordant lactate_2 (lac_2) results were obtained on multiple patient samples on an atellica ch 930 analyzer.The results were reported to the physician(s) and at least three results were questioned.These three patients were redrawn and the new samples were processed on an alternate atellica ch 930 analyzer, resulting higher.The customer is unable to provide whether additional patients will be redrawn and retested and did not quantify the number of samples impacted by this issue.This mdr is being filed in an abundance of caution as the customer did not provide details of all affected samples.There are no known reports of patient intervention or adverse health consequences due to the discordant lac_2 results.
|
| |
|
Additional Manufacturer Narrative
|
|
A united states customer contacted a siemens customer care center to report discordant lactate_2 (lac_2) results were obtained on multiple patient samples on an atellica ch 930 analyzer.The customer determined that these patient results were generated from a reagent pack that was incorrectly prepared by the customer.Additionally, quality controls was out of range on the reagent pack.Siemens further investigated the issue and determined the probable cause of the discordant patient results was due to preanalytical error, in which the customer did not follow the atellica ch lactate_2 (lac_2) instructions for use for the proper preparation of the reagents.The cause of the discordant lac_2 results is due to user error.This mdr is being filed in an abundance of caution.The instrument and reagent are performing to specifications.No further evaluation of this device is required.
|
| |
|
Search Alerts/Recalls
|