• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY SYNCHRO SELECT-14 SUPPORT ST 300CM; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY SYNCHRO SELECT-14 SUPPORT ST 300CM; WIRE, GUIDE, CATHETER Back to Search Results
Model Number SSUP300STR
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is not available to the manufacturer.
 
Event or Problem Description
It was reported that a vascular dissociation procedure was performed in a patient in intracranial blood vessel to place a stent.Initially during the procedure, resistance was felt when the physician advanced the subject guidewire through the balloon catheter for pre-extension.Later when the stent system was advanced over the subject guidewire, the resistance was so strong that it was determined that the subject guidewire cannot be used anymore because of the positioning accuracy.According to the physician the subject guidewire might have had the coating problem.The subject guidewire was replaced and the procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Additional Manufacturer Narrative
D4 expiration date - added.H4 manufacturing date ¿ added.There are controls in the manufacturing process to ensure the product met specifications upon release.The subject device is not available; therefore, functional testing as well as visual testing cannot be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.Additional information provided by the customer indicated that the device was confirmed to be in good condition during preparation/prior to use on the patient and was prepared as per the dfu, continuous flush was set up and maintained throughout the clinical procedure, the tortuosity of the patient's anatomy was normal, the provided introducer was used to load the guidewire, the guidewire was shaped 45 degrees, and it was unknown where on the guidewire the coating issue was noted.The same balloon catheter and stent were used to complete the procedure after replacing the subject device with another.While there are a number of potential causes for the reported 'guidewire difficult to advance' and 'guidewire delamination', because review and analysis of available information failed to identify a definitive cause and the device was not returned, an assignable cause of undeterminable was assigned to this complaint.
 
Event or Problem Description
It was reported that a vascular dissociation procedure was performed in a patient in intracranial blood vessel to place a stent.Initially during the procedure, resistance was felt when the physician advanced the subject guidewire through the balloon catheter for pre-extension.Later when the stent system was advanced over the subject guidewire, the resistance was so strong that it was determined that the subject guidewire cannot be used anymore because of the positioning accuracy.According to the physician the subject guidewire might have had the coating problem.The subject guidewire was replaced and the procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHRO SELECT-14 SUPPORT ST 300CM
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY
4870 west 2100 south
salt lake city UT 84120
Manufacturer (Section G)
STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY
4870 west 2100 south
salt lake city UT 84120
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key13608648
Report Number3012931345-2022-00035
Device Sequence Number11245501
Product Code MOF
UDI-Device Identifier07613327508628
UDI-Public07613327508628
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K202522
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/03/2024
Device Model NumberSSUP300STR
Device Catalogue NumberSSUP300STR
Device Lot Number0000063386
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/03/2022
Supplement Date Received by Manufacturer04/21/2022
Initial Report FDA Received Date02/26/2022
Supplement Report FDA Received Date05/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/05/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GATEWAY MR BALLOON (STRYKER); WINGSPAN STENT (STRYKER)
Patient SexUnknown
-
-