Model Number 286760350
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/27/2022
Type of Reportable Event Malfunction
Event or Problem Description
Device report from (b)(6) reports an event as follows: it was reported that during a tlif (s1) for treating lumbar spinal canal stenosis on (b)(6) 2022, the ring on the sleeve came off during sleeve turning for screw deployment.The procedure was completed with no delay.Patient was stable.This report is for a viper insertion sleeve.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Additional narrative: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: a product investigation was conducted.The product was returned to us cq for evaluation.The us cq team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the x-tab insertion sleeve was missing its color marker component.No signs of any breakages could be observed.No other issues were identified.The dimensional inspection was not performed as there was definitive evidence of a missing component.The observed condition of the x-tab insertion sleeve was consistent with a random component failure that may have been caused by exposure to excessive/unintended forces.A functional test could not be performed as no mating devices were returned for evaluation.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the x-tab insertion sleeve was observed to have a missing color marker component.While no definitive root cause could be determined from the available information, it is probable that the color marker component of the x-tab insertion sleeve was missing due to the exposure of the device to excessive/unintended forces.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.A review of the receiving inspection (ri) for x-tab insertion sleeve was conducted identifying that lot number ag2127361 was released in one batch.Batch1: lot qty of 82 units were released on 06 aug 2013 with no discrepancies.As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X-TAB INSERTION SLEEVE
Common Device Name
CANNULA, SURGICAL, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key13608656
Report Number1526439-2022-00290
Device Sequence Number1930975
Product Code GEA
UDI-Device Identifier10705034214955
UDI-Public(01)10705034214955
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number286760350
Device Catalogue Number286760350
Device Lot NumberAG2127361
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/23/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/27/2022
Supplement Date Received by Manufacturer04/21/2022
Initial Report FDA Received Date02/26/2022
Supplement Report FDA Received Date04/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/06/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN SCREWS
Patient SexUnknown
-
-