• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL FAST-CATH¿ DUO HEMOSTASIS INTRODUCER, CATH-LOCK¿ LOCKING HUB, 23 CM SHEATH, 9 F; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL FAST-CATH¿ DUO HEMOSTASIS INTRODUCER, CATH-LOCK¿ LOCKING HUB, 23 CM SHEATH, 9 F; INTRODUCER, CATHETER Back to Search Results
Model Number 406333
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Based on the information received, the cause of the reported leak remains unknown.The ifu states damage to the valve assembly may occur if inner catheter is withdrawn rapidly.
 
Event or Problem Description
Related manufacturer ref# 3005334138-2022-00073.During a paroxysmal supraventricular tachycardia ablation procedure, a blood was noted from the hemostasis valves of the sheath.The sheath set was replaced which did not resolve the issue.The sheath was replaced and the procedure was completed with no adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FAST-CATH¿ DUO HEMOSTASIS INTRODUCER, CATH-LOCK¿ LOCKING HUB, 23 CM SHEATH, 9 F
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key13616360
Report Number3005334138-2022-00072
Device Sequence Number19376795
Product Code DYB
UDI-Device Identifier05414734203876
UDI-Public05414734203876
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K061015
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2022
Device Model Number406333
Device Catalogue Number406333
Device Lot Number7055582
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/16/2022
Initial Report FDA Received Date02/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/20/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-