MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751LNAS; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-751LNAS
Device Problem Failure to Cycle (1142)
Patient Problem Hyperglycemia (1905)
Event Date 02/24/2022
Event Type  Injury  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that they were experienced high blood glucose level.The customer¿s blood glucose level was 400 mg/dl at the time of incident.The customer stated that insulin pump had compromised force sensor system alarm and drive support cap was normal.The customer will continue to use the device.
 
Manufacturer Narrative
Unit received motor error alarm during basic occlusion test due to faulty fsr (gold).Unable to perform basic occlusion test, occlusion test, prime/a33 test, excessive no delivery test or verify a33 alarm due to motor error alarm.However, unit passed rewind test and displacement test.Pump history download using thds was successful.However, a47 alarms found on the history file which caused corrupted data.Unable to confirm the reported event date due to the corrupted data.A47 alarms are due to the pump not having power for a long period of time.No moisture damage found on the electronics, motor, battery tube and vibrator assembly noted.Motor passed motor test.The test p-cap/reservoir does lock into place.Unit had scratched case, cracked battery tube threads, cracked reservoir tube lip, cracked reservoir tube window, cracked case at reservoir tube window corners, stained end cap sticker, stained address/serial number label and scratched keypad overlay.Unable to confirm a33 alarm due to motor error alarm.Motor error alarm during basic occlusion test due to faulty fsr (gold) confirmed.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Manufacturer Narrative
Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.The information that provided with the initial report was incomplete.The complete information has been included with this report in b5 section.The information related to health effect- impact code and component code has been updated and provided in h6 section of this report.
 
Event Description
It was reported that the customer discontinued the use of insulin pump and return insulin pump for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-751LNAS
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
tricha miles
ceiba norte ind. park #50 road
juncos 00777--386
7635140379
MDR Report Key13617888
MDR Text Key286228246
Report Number2032227-2022-118854
Device Sequence Number1
Product Code OZO
UDI-Device Identifier000000643169933224
UDI-Public(01)000000643169933224
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 12/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/28/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-751LNAS
Device Catalogue NumberMMT-751LNAS
Device Lot NumberA5751LNASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received12/18/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/06/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
FRN-UNK-RSVR, UNOMED INF SET
Patient Outcome(s) Other;
Patient Age50 YR
Patient SexMale
Patient Weight88 KG
-
-