• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON LABORATORIES IRELAND LTD. ACRYSOF IQ NATURAL SINGLEPIECE IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LABORATORIES IRELAND LTD. ACRYSOF IQ NATURAL SINGLEPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number SN60WF
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/18/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A healthcare professional reported during an intraocular lens implant procedure, the operating team had to use a backup lens since the regular lens was found out to be deformed.Additional information has been requested.
 
Additional Manufacturer Narrative
Based on information received following submission of the initial report, this event does not meet criteria for reporting as a device malfunction.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
Additional information has been requested and received which state there was no patient contact.
 
Additional Manufacturer Narrative
The product was returned for analysis and the reported damage was observed.Lens returned pressed against one(1) post of the lens case base resulting in deformation of the iol.A significant amount of solution is dried on the iol.Both haptics are bent/deformed.The optic is bent/deformed and the optic edge is chipped.We are unable to determine the root cause for the reported complaint - "deformed iol".The returned iol shows evidence of possible handling by the customer due to the presence of solution dried on the iol.In addition to this, all iols are 100% cosmetically inspected as per approved manufacturing procedures and the observed lens damage would not meet our current release criteria.Based on these investigation findings, we are unable to verify if the iol contributed to the event.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ NATURAL SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key13618122
Report Number9612169-2022-00089
Device Sequence Number3927332
Product Code HQL
UDI-Device Identifier00380655093283
UDI-Public00380655093283
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2025
Device Model NumberSN60WF
Device Catalogue NumberSN60WF.245
Device Lot Number21285129
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/03/2022
Supplement Date Received by Manufacturer04/12/2022
04/20/2022
Initial Report FDA Received Date02/28/2022
Supplement Report FDA Received Date05/09/2022
05/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DISCOVISC OPHTHALMIC VISCOSURGICAL DEVICE; MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D
Patient Age62 YR
Patient SexFemale
-
-