| Model Number |
A98264 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
02/10/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The full patient identifier is (b)(4).The customer did not supply patient demographics such as age, date of birth, sex, weight, ethnicity or race.A beckman coulter local support assisted the customer with verifying the system performance.From (b)(6) to (b)(6) 2022, the customer ran test samples in duplicate as part of troubleshooting.The customer provided six (6) test samples results using reagent pack serial number (b)(4) and (b)(4), which were also non-reproducible: test sample one : initial result 155.3 pg/ml repeated at 6.6 pg/ml.Test sample two : initial result 2262.2 pg/ml repeated at 5.2 pg/ml.Test sample three : initial result 1764.4 pg/ml repeated at 2.5 pg/ml.Test sample four : initial result 972.6 pg/ml repeated at 37.78 pg/ml.Test sample five : initial result 47.71 pg/ml repeated at 9.75 pg/ml.Test sample six : initial result 70.97 pg/ml repeated at 21.03 pg/ml.The non-reproducible tnidx recovery was not observed in quality control samples run on pack serial number (b)(4) or other serial numbers.Running tnidx qc (quality control) recovery with consistent results obtained was reported to recover at <6 %cv (percent coefficient of variation).System performance indicators of high sensitivity system check, system check, carryover and tnidx precision studies (using reagent pack serial numbers (b)(4) and (b)(4)) were run on (b)(6) 2022.All passed within beckman coulter specifications and did not demonstrate an issue with the system.Additional testing on the reagent pack serial numbers (b)(4) and (b)(4) could not be performed as no remaining product was available for testing.Per communication with local support, no abnormalities were observed with reagent packs.Two retain packs of tnia2 lot 124617 were visually inspected by complaint investigator on (b)(6) 2022.The packs were unremarkable upon visual inspection.No precipitate was observed in the retain reagent packs.On (b)(6) 2022, the most recent three months ((b)(6) 2021, (b)(6) 2021, (b)(6) 2021) of completed immunoassay complaint handling unit trending workbooks were reviewed by complaint investigator.No complaint trending triggers were identified for accutni+3 reagent lot a98264 involving reagent lot number 124617.In conclusion, an assignable cause of the event cannot be determined with the available information.There is insufficient evidence to suggest a reagent or hardware malfunction occurred in this event.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2022 the customer reported obtaining one non-reproducible falsely elevated troponin i (access accutni+3) result for one patient involving the laboratory's unicel dxi 800 access immunoassay analyzer (serial number (b)(4)).On (b)(6) 2022 the initial tnidx patient result was 838.8 pg/ml.The sample was repeated twice with the following lower results: 4.0 pg/ml and 3.7 pg/ml.All three sample replicates were run on the reagent pack lot 124617, pack serial number (b)(4).The customer reported the patient received cardiac catheter ablation treatment due to the falsely elevated tnidx result.Beckman coulter local support assisted the customer with verifying the system performance.From (b)(6) to (b)(6) 2022, the customer ran test samples in duplicate as part of troubleshooting.The customer provided six (6) test samples results using reagent pack serial number 499 and 7562, which were also non-reproducible: test sample one: initial result 155.3 pg/ml repeated at 6.6 pg/ml.Test sample two: initial result 2262.2 pg/ml repeated at 5.2 pg/ml.Test sample three: initial result 1764.4 pg/ml repeated at 2.5 pg/ml.Test sample four: initial result 972.6 pg/ml repeated at 37.78 pg/ml.Test sample five : initial result 47.71 pg/ml repeated at 9.75 pg/ml.Test sample six : initial result 70.97 pg/ml repeated at 21.03 pg/ml.The non-reproducible tnidx recovery was not observed in quality control (qc) samples.Tnidx qc was run across multiple reagent pack serial numbers with consistent results obtained.Running qc recovery was reported to recover at <6 %cv (percent coefficient of variation).System performance indicators of high sensitivity system check, system check, carryover and tnidx precision studies (using reagent pack serial numbers (b)(4) and (b)(4)) were run on (b)(6) 2022.All passed within beckman coulter specifications and did not demonstrate an issue with the system.Per communication with local support, no abnormalities were observed with reagent packs.The patient sample was serum.There were no issues with sample integrity reported by the customer, no fibrin or particulate matter was observed.Sample information such as sample collection tube used, centrifugation time and speed, storage or handling was not provided by the customer.No further sample information was provided.
|
| |
|
Search Alerts/Recalls
|
|
|