• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. RINGLOC BI-POLAR 28X44MM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. RINGLOC BI-POLAR 28X44MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Codes Off-Label Use (1494); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 11/24/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Concomitant products: 110010273- g7 osseoti multihole 70mm i-3586254 ; 00625006530-trilogy bone scr 6.5x30-j6979290; 00625006520-trilogy bone scr 6.5x20-j6864845; 30154409-44mm i.D.Size i +5mm offset liner-64768030.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2021 - 03457.Customer has indicated that the product will not be returned to zimmer biomet for investigation, product location unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Device location unknown.
 
Event or Problem Description
It was reported patient underwent a left hip revision approximately 2 months¿ post implantation due to implant loosening.During the procedure all implants were removed including a 44 ringloc bipolar with a 44 +5 liner which were off-label use, surgeon was aware.The patient now only has cement in the hip area.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional information on the reported event.
 
Event or Problem Description
No additional information on the reported event.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: g3; h2; h3; h6 reported event was unable to be confirmed.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.A review of the device history records identified no related deviations or anomalies during manufacturing.A definitive root cause cannot be determined however it was noted that the combination of products used together is considered off-label and may have caused or contributed to the event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RINGLOC BI-POLAR 28X44MM
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13618632
Report Number0001825034-2022-00381
Device Sequence Number17553702
Product Code JDI
UDI-Device Identifier00880304209640
UDI-Public(01)00880304209640(17)260623(10)936010
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K051569
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/23/2026
Device Model NumberN/A
Device Catalogue Number11-165212
Device Lot Number936010
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/02/2021
Supplement Date Received by Manufacturer02/28/2022
03/15/2022
Initial Report FDA Received Date02/28/2022
Supplement Report FDA Received Date03/02/2022
03/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/23/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
NI
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-