Model Number IBI-81594
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/18/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
One decapolar, spiral loop, inquiry afocus ii diagnostic catheter was received for evaluation.Electrode 3 was missing from the catheter.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the missing electrode ring remains unknown.
 
Event or Problem Description
This report is to address the investigation findings of a missing electrode on the catheter tip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INQUIRY¿ AFOCUS LL¿ DIAGNOSTIC CATHETER
Common Device Name
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
2375 morse ave
irvine CA 92614
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
2375 morse ave
irvine CA 92614
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key13618817
Report Number2030404-2022-00015
Device Sequence Number17596622
Product Code DRF
UDI-Device Identifier05414734305006
UDI-Public05414734305006
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K010471
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Model NumberIBI-81594
Device Catalogue NumberIBI-81594
Device Lot Number7914678
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/09/2022
Initial Report FDA Received Date02/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-