| Model Number |
SHORT-HD |
| Medical Device Problem Code |
Excessive Heating (4030)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by (b)(6) that during service and evaluation, it was determined that the attachment device had a worn bearing, produced heat and had an illegible labelling etch.It was further determined that the device failed pretest for labeling assessment and verification: temperature.It was noted that the tringle icon was missing and an additional label was added to the extension sleeve by the customer.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2022.All available information has been disclosed.If additional information should available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed, and it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to component failure from wear.Udi: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed, and it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to component failure from wear.Udi: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported by (b)(6) that during service and evaluation, it was determined that the attachment device had a worn bearing, produced heat and had an illegible labelling etch.It was further determined that the device failed pretest for labeling assessment and verification: temperature.It was noted that the tringle icon was missing and an additional label was added to the extension sleeve by the customer.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2022.All available information has been disclosed.If additional information should available, a supplemental medwatch will be submitted accordingly.
|
| |
|