MAUDE Adverse Event Report: STRYKER TRAUMA KIEL TROCHANTERIC NAIL KIT, TI GAMMA3® Ø11X200MM X 130°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL TROCHANTERIC NAIL KIT, TI GAMMA3® Ø11X200MM X 130°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 3130-1200S
Medical Device Problem Code Device Difficult to Setup or Prepare (1487)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/02/2022
Type of Reportable Event Malfunction
Event or Problem Description
Excessive force was used for the insertion of the cephalic screw, it was noticeable that it did not slide well inside the nail.It was checked that everything was well connected.Nail and screw was disassembled and reassembled outside, it continued to rub against the cephalic screw when passing inside of the nail.A new nail and screw was used at the end.There was a delay of 1 hour during the procedure.
 
Additional Manufacturer Narrative
Upon completion of investigation, additional information will be provided in a supplemental report.
 
Additional Manufacturer Narrative
The reported event could not be confirmed, since the returned device was found to be fully functional and conforming to specifications.The device inspection revealed the following the received device overall appears to be in an alright state.With a closer look at the lag screw hole, heavy stress marks are visible inside the bore which was caused by the use of excessive force during the insertion of the lag screw.Functional test was performed with the reported lag screw and nail, in which no insertion problem was observed, and the lag screw could be locked as required.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on the investigation, the root cause was attributed to a user related issue since the device is found to be fully functional during the inspection.If any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
Excessive force was used for the insertion of the cephalic screw, it was noticeable that it did not slide well inside the nail.It was checked that everything was well connected.Nail and screw was disassembled and reassembled outside, it continued to rub against the cephalic screw when passing inside of the nail.A new nail and screw was used at the end.There was a delay of 1 hour during the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TROCHANTERIC NAIL KIT, TI GAMMA3® Ø11X200MM X 130°
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13619684
Report Number0009610622-2022-00074
Device Sequence Number4673901
Product Code HSB
UDI-Device Identifier04546540375131
UDI-Public04546540375131
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K034002
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2026
Device Model Number3130-1200S
Device Catalogue Number31301200S
Device Lot NumberK05CC52
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/03/2022
Supplement Date Received by Manufacturer05/19/2022
Initial Report FDA Received Date02/28/2022
Supplement Report FDA Received Date06/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-