• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ERIKA DE REYNOSA, S.A. DE C.V. SAFE LOCK APD LUER LOCK CONNECTOR; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ERIKA DE REYNOSA, S.A. DE C.V. SAFE LOCK APD LUER LOCK CONNECTOR; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE Back to Search Results
Model Number 050-95018
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/21/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a peritoneal dialysis (pd) patient discovered a fluid leak coming from the safe lock connector that was connected to the baxter bag during their pd treatment.The patient reported receiving an air detected in cassette alarm during dwell 3 of 4 of treatment and the treatment was cancelled.The cause of the leak is unknown.Fluid did not come in contact with the cycler.It was reported that an alternate treatment option was available.Upon follow up, the patient confirmed being trained on performing stat drains, as well as manual peritoneal dialysis therapy if needed.The patient did not develop any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The patient is continuing peritoneal dialysis therapy with no further issues.The adaptor used by the patient was discarded and is not available for return for physical evaluation by the manufacturer.
 
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Additional Manufacturer Narrative
Plant investigation: the sample was not returned to the manufacturer and the lot number was not provided.A manufacturing review was performed on the products shipped to the patient for the three (3) month time frame which immediately preceded the event occurrence date.This review included the lot numbers for all fresenius safe lock apd luer lock connector shipped to this account within the selected time frame.The entire set of lots have been sold and distributed.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.An investigation of the device history records (dhr) was conducted and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The product lots involved met all specifications for release.A review of the dhr did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
 
Event or Problem Description
It was reported that a peritoneal dialysis (pd) patient discovered a fluid leak coming from the safe lock connector that was connected to the baxter bag during their pd treatment.The patient reported receiving an air detected in cassette alarm during dwell 3 of 4 of treatment and the treatment was cancelled.The cause of the leak is unknown.Fluid did not come in contact with the cycler.It was reported that an alternate treatment option was available.Upon follow up, the patient confirmed being trained on performing stat drains, as well as manual peritoneal dialysis therapy if needed.The patient did not develop any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The patient is continuing peritoneal dialysis therapy with no further issues.The adaptor used by the patient was discarded and is not available for return for physical evaluation by the manufacturer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFE LOCK APD LUER LOCK CONNECTOR
Common Device Name
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
director, quality systems
1100 e, military hwy, suite c
pharr TX 78577
Manufacturer Contact
jason busch
920 winter st
waltham, MA 02451
9043166958
MDR Report Key13619850
Report Number8030665-2022-00193
Device Sequence Number4368800
Product Code KDJ
UDI-Device Identifier00840861100811
UDI-Public00840861100811
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K124024
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number050-95018
Device Catalogue Number050-95018
Was Device Available for Evaluation? No
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2022
Supplement Date Received by Manufacturer03/11/2022
Initial Report FDA Received Date02/28/2022
Supplement Report FDA Received Date03/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DELFLEX PD FLUID; DELFLEX PD FLUID; LIBERTY CYCLER SET ; LIBERTY CYCLER SET ; LIBERTY SELECT CYCLER; LIBERTY SELECT CYCLER
Patient Age86 YR
Patient SexMale
Patient Weight86 KG
-
-