• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. 36MM I.D. SIZE LL NEUTRAL LINER USE WITH 58MM O.D. SIZE LL SHELL; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. 36MM I.D. SIZE LL NEUTRAL LINER USE WITH 58MM O.D. SIZE LL SHELL; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/04/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the poly liner did not seat correctly into the cup.Cup was able to seat correctly with new liner.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
(b)(4).Concomitant medical products: 00875705801-58mm o.D.Size ll porous uncemented with cluster holes shell use with ll liners-64904566.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: g3; h2; h3; h6 complaint sample was evaluated and the reported event was confirmed.Visual review of the device confirms item and lot etch.Outer spherical surface shows screw hole damage.Side view of locking feature, anti-rotation scallops, and face of device shows nicks, gouges, and deformation from attempted implant and explant.Screw deformed inner and outer spherical surface.No other damage was noted.Review of the device history records identified no (related) deviations or anomalies during manufacturing.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional information on the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
36MM I.D. SIZE LL NEUTRAL LINER USE WITH 58MM O.D. SIZE LL SHELL
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13620293
Report Number0001822565-2022-00626
Device Sequence Number14621072
Product Code JDI
UDI-Device Identifier00889024149236
UDI-Public(01)00889024149236(17)220630(10)63672995
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K200823
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2022
Device Model NumberN/A
Device Catalogue Number00875101336
Device Lot Number63672995
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/04/2022
Supplement Date Received by Manufacturer05/03/2022
Initial Report FDA Received Date02/28/2022
Supplement Report FDA Received Date05/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/23/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
NI
Patient Sequence Number1
Patient SexMale
-
-