• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS AG TRIAL FEMORAL HEAD 36 XS/-3; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS AG TRIAL FEMORAL HEAD 36 XS/-3; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS Back to Search Results
Model Number 75100855
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/08/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during instrument inspection on facility, the tine inside of the trial femoral head 36 xs/-3 broke.No case involved.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
H3, h6: it was reported that during instrument inspection on facility, the tine inside of the trial femoral head 36 xs/-3 broke.The device, intended for use in treatment, was not returned for investigation.A product evaluation could therefore not be performed.A review of the production documentation did not detect any deviation that could have contributed to the reported failure mode.A review of the complaint history of the affected device was performed.Only one additional complaint for the batch in question was detected, but with another reported failure mode.The failure mode and the severity are covered in the corresponding risk management document.There is no indication that the reported device failed to meet manufacturing specifications upon release for distribution.A review of past corrective actions was performed.Previous investigations demonstrated that the trial femoral head may disconnect from the stem taper intra-operatively and flaps might break off.These issues are due to high loads, beyond the design limit, that may arise on the tabs during manipulation.The failure modes may further be exacerbated by excessive use of the device over time.The performance of the device is within the risks level that are anticipated in the risk management documentation.The current design remains therefore state of the art.As part of the continuous improvement smith + nephew is nonetheless working on a design enhancement, options are currently being reviewed and tested at our facility.The current design will be further monitored through post-market surveillance processes.No further actions are necessary at this time point.Based on the clear complaint description, the reported failure mode of a flap breakage can be confirmed.The root cause is attributed to an insufficient design specification.According to document "processing (cleaning, disinfection and sterilization) of instruments from smith & nephew orthopaedics ag" (lit.N°03389-en 1363 v3 11/19), all devices must be inspected and controlled for proper functioning after cleaning/disinfection.Should the part be returned or should more information become available, the investigation will be reopened.No further actions are deemed necessary at the time.Smith+nephew will continue to monitor this device for similar issues.
 
Additional Manufacturer Narrative
H6: it was reported that during instrument inspection on facility, the tine inside of the trial femoral head 36 xs/-3 broke.The device, intended for use in treatment, was not returned for investigation.A product evaluation could therefore not be performed.A review of the production documentation did not detect any deviation that could have contributed to the reported failure mode.A review of the complaint history of the affected device was performed.Only one additional complaint for the batch in question was detected, but with another reported failure mode.The failure mode and the severity are covered in the corresponding risk management document.There is no indication that the reported device failed to meet manufacturing specifications upon release for distribution.A review of past corrective actions was performed.Previous investigations demonstrated that the trial femoral head may disconnect from the stem taper intra-operatively and flaps might break off.These issues are due to high loads, beyond the design limit, that may arise on the tabs during manipulation.The failure modes may further be exacerbated by excessive use of the device over time.The performance of the device is within the risks level that are anticipated in the risk management documentation.The current design remains therefore state of the art.As part of the continuous improvement smith + nephew is nonetheless working on a design enhancement, options are currently being reviewed and tested at our facility.The current design will be further monitored through post-market surveillance processes.No further actions are necessary at this time point.Based on the clear complaint description, the reported failure mode of a flap breakage can be confirmed.The root cause is attributed to an insufficient design specification.According to document "processing (cleaning, disinfection and sterilization) of instruments from smith & nephew orthopaedics ag" (lit.N°03389-en 1363 v3 11/19), all devices must be inspected and controlled for proper functioning after cleaning/disinfection.Should the part be returned or should more information become available, the investigation will be reopened.No further actions are deemed necessary at the time.Smith+nephew will continue to monitor this device for similar issues.Case-(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIAL FEMORAL HEAD 36 XS/-3
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ   CH-5000
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13620915
Report Number9613369-2022-00039
Device Sequence Number4410009
Product Code LZO
UDI-Device Identifier07611996121889
UDI-Public07611996121889
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K132435
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number75100855
Device Catalogue Number75100855
Device Lot NumberA60203
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/08/2022
Supplement Date Received by Manufacturer03/18/2022
03/18/2022
Initial Report FDA Received Date02/28/2022
Supplement Report FDA Received Date03/20/2022
03/21/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/05/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-