ABBOTT VASCULAR G4 STEERABLE GUIDING CATHETER; CATHETER, STEERABLE Back to Search Results
Catalog Number SGC0702
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Thrombosis/Thrombus (4440); Unspecified Tissue Injury (4559)
Event Date 02/08/2022
Event Type  Injury  
Event Description
This is filed to report thrombus and tissue damage.It was reported this was a mitraclip procedure to treat functional mitral regurgitation (mr) with a grade of 3.Two clips were deployed on the mitral valve, reducing mr to a grade of <1.The steerable guide catheter (sgc) was retracted from the left atrium to the right atrium, but a fluttering object was observed at the septum.It was unknown if the object was thrombus or tissue.Aspiration was performed and the object was successfully removed.There was no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
The device will not be returned for evaluation, the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Based on available information, the cause of the reported thrombus and unspecified tissue injury could not be determined.Thrombosis/thrombus and tissue injury are listed in the mitraclip system instructions for use (ifu) as known possible complications associated with mitraclip procedures.The reported unexpected medical intervention was a result of case-specific circumstances.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G4 STEERABLE GUIDING CATHETER
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13620925
MDR Text Key286583816
Report Number2024168-2022-02061
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K190167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/28/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/20/2022
Device Catalogue NumberSGC0702
Device Lot Number10921R249
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/09/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/21/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
TWO IMPLANTED MITRACLIP'S
Patient Outcome(s) Required Intervention;
-
-