Model Number 357.515
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on (b)(6) 2022, one of the sterile processing technicians was putting a set back together after use and noticed the silver pin on the brown handle of the 8mm ball hex screwdriver was sitting proud.It is unknown how the surgery was completed.Patient outcome is unknown.This complaint involves one (1) device ball hex scrdriver 8 for ti fem nail.This is report 1 of 1 for complaint (b)(4).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Additional narrative: complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.A review of the device history record has been completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h6: a product investigation was conducted.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the silver pin is slightly protruded from the screwdriver handle of the ball hex scrdriver 8 for ti fem nail, this condition is related with the customer complaint.Drawing/specifications review: based on the date of manufacture, the current and manufactured revision of drawings were reviewed.No design issues or discrepancies were identified.A dimensional inspection was performed to the screwdriver handle and the pin due to the protruded condition noted on the pin.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces during the use of the device.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the silver pin was noted protruded from the screwdriver handle of the ball hex scrdriver 8 for ti fem nail.No definitive root cause could be determined.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.H3, h6: a device history record (dhr) review was conducted: part # 357.515.Synthes lot # 5120403.Supplier lot # n/a.Release to warehouse date: 02 dec2005.Manufactured by: synthes brandywine.No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BALL HEX SCRDRIVER 8 FOR TI FEM NAIL
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
BRANDYWINE
1303 goshen parkway
west chester PA 19380
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key13621076
Report Number2939274-2022-00692
Device Sequence Number4368750
Product Code HXX
UDI-Device Identifier10886982196514
UDI-Public(01)10886982196514
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number357.515
Device Catalogue Number357.515
Device Lot Number5120403
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/31/2022
Supplement Date Received by Manufacturer03/29/2022
Initial Report FDA Received Date02/28/2022
Supplement Report FDA Received Date03/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/02/2005
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-