• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. BALLARD CS NEO/PED, 6 FR, Y-ADAPTER; VAP CLOSED SUCTION CATHETERS & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. BALLARD CS NEO/PED, 6 FR, Y-ADAPTER; VAP CLOSED SUCTION CATHETERS & ACCESSORIES Back to Search Results
Model Number 196
Medical Device Problem Code Suction Problem (2170)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The sample is reported to be available, but has not yet been received by the manufacturer.A review of the device history record is in-progress.All information reasonably known as of 24-feb-2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).
 
Event or Problem Description
It was reported the "ballard catheter was found to be defective on a patient." there was no reported injury.The suction thumb knob would not come back to the normal position after depressing for suction.Additional information received 02-feb-2022 stated the device was in use for 5-days prior to the reported incident occurring.
 
Additional Manufacturer Narrative
The device history record for the reported lot number, 30151032, was reviewed and documented that the lot was manufactured according to the approved manufacturing procedures, specifications, and then released by quality assurance.All information reasonably known as of 29-apr-2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BALLARD CS NEO/PED, 6 FR, Y-ADAPTER
Common Device Name
VAP CLOSED SUCTION CATHETERS & ACCESSORIES
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT S DE RL DE CV
carretera internacional
salida norte no. 1053
magdalena, cp
Manufacturer Contact
lisa clark
5405 windward parkway
alpharetta, GA 30004
4704485444
MDR Report Key13621951
Report Number8030647-2022-00067
Device Sequence Number12026850
Product Code BSY
UDI-Device Identifier00609038938240
UDI-Public00609038938240
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/19/2026
Device Model Number196
Device Catalogue NumberN/A
Device Lot Number30151032
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/01/2022
Supplement Date Received by Manufacturer04/02/2022
Initial Report FDA Received Date02/28/2022
Supplement Report FDA Received Date04/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-