Model Number 0684-00-0480-02
Medical Device Problem Code Device Markings/Labelling Problem (2911)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complete initial reporter name: (b)(6).Event site postal code: (b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
 
Event or Problem Description
It was reported that the intra-aortic balloon (iab) label sticker had two different sizes mentioned on it.The customer stated that there were three iabs with this same issue.There was no patient involvement and no adverse event reported.This report is for one of the three iabs.Two additional reports will be submitted for the other two iabs.
 
Additional Manufacturer Narrative
The product was not returned for evaluation.Photos were provided and reviewed.Inspection of the provided photos determined that the labeling was incorrect.The evaluation confirms the reported problem.We are unable to determine how this failure may have occurred and has been escalated to capa and a field action has been initiated.A non-conforming material report review was completed for the reported product.No nonconformances were found that are considered to be related to the event.Reference complaint #: (b)(4).
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINEAR 7.5 FR. 40CC IAB
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer (Section G)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer Contact
brian schaeffer
15 law drive
fairfield, NJ 
MDR Report Key13623130
Report Number2248146-2022-00163
Device Sequence Number14769718
Product Code DSP
UDI-Device Identifier10607567106571
UDI-Public10607567106571
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K041281
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 08/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/18/2023
Device Model Number0684-00-0480-02
Device Catalogue Number0684-00-0475
Device Lot Number3000131710
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2022
Supplement Date Received by Manufacturer08/24/2022
Initial Report FDA Received Date02/28/2022
Supplement Report FDA Received Date08/24/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/18/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-