• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. NV GXL LNR, +5LAT, 36MM G3-56/58MM CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. NV GXL LNR, +5LAT, 36MM G3-56/58MM CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number NV GXL LNR, +5LAT, 36MM G3-56/58MM CUPS
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 02/07/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Event or Problem Description
As reported, this patient had a revision due to premature polyethylene wear and osteolysis.No additional information available.
 
Additional Manufacturer Narrative
Section b5: additional information received indicates that patient revision was performed on (b)(6)2022.Patient was revised with a competitors.The stem was retained.Patient is a male, 70 years old.Section d10: concomitant medical products: element-stem, collared w/ha, std offset, sz 11 (cat# 164-03-11 / serial# (b)(6)).Biolox delta femoral head 36mm od, +0mm (cat# 170-36-00 / serial# (b)(6)).Nv crown cup clstr hole 58mm group 3 (cat# 180-01-58 / serial# (b)(6)).Additional information, including the product investigation, will be submitted within 30 days of receipt section d2: product code.
 
Additional Manufacturer Narrative
Section h10: (h3) based on the available information, the patient involved meets the following risk criteria for early prosthesis wear/osteolysis as specified in the hhe: implanted with a lateralized liner and implanted with a component having a shelf age of greater than 2 years.The most likely cause for the revision reported due to early prosthesis wear and osteolysis is a combination of the risk factors specified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NV GXL LNR, +5LAT, 36MM G3-56/58MM CUPS
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66 court
gainesville, FL 32653
3523771140
MDR Report Key13623208
Report Number1038671-2022-00224
Device Sequence Number12025305
Product Code JDI
UDI-Device Identifier10885862024336
UDI-Public10885862024336
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K070479
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/11/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/27/2014
Device Model NumberNV GXL LNR, +5LAT, 36MM G3-56/58MM CUPS
Device Catalogue Number136-36-53
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2022
Supplement Date Received by Manufacturer03/15/2022
06/10/2022
Initial Report FDA Received Date02/28/2022
Supplement Report FDA Received Date03/31/2022
07/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/29/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient SexMale
-
-